The user manual of KC-programmable salt spray tester describes the scope of application, test method, test liquid preparation and precautions of salt spray test (Chinese standard).
This manual provides instructions for the Kecheng KC- programmable salt spray tester, produced by Ningbo Beilun District Kecheng Instrument Co., Ltd. It details the equipment's construction and installation requirements, ensuring the use of inert materials and proper air filtration to prevent corrosion and contamination during testing.
The document outlines standard operating procedures for salt spray testing according to Chinese National Standards. It specifies critical test conditions, including sodium chloride concentration (40~60 g/L), pH levels (6.5-7.2), and temperature controls for the pressure chamber (47±1°C) and salt water chamber (35±1°C). Additionally, it covers sample placement, post-test handling, and maintenance guidelines. This manual is written in Chinese.
| Spray Rate | 1.0~2.0 ml/80cm2/h |
|---|---|
| Pressure Chamber Temperature | 47±1 ℃ |
| Salt Water Chamber Temperature | 35±1 ℃ |
| Laboratory Relative Humidity | 85% or above |
| Sample Dimensions | 150×70mm or 100×65mm |
The concentration of the sodium chloride solution must be maintained between 40 and 60 g/L during the test.
The temperature of the pressure chamber must be kept at 47±1 ℃.
The main surface of the sample should be inclined at 15 to 30 degrees to the vertical line and parallel to the main flow direction of the spray.