CC2674-10 type high voltage withstand voltage tester is a high voltage withstand voltage tester widely used in various special electrical equipment withstand voltage tests, as well as necessary equipment for relevant laboratories and inspection rooms.
Changchuang CC2674-10 is a super high-voltage withstand voltage tester used for withstand testing of special electrical appliances and in laboratories and inspection rooms. The manual describes its specifications, precautions, front-panel functions, operating method, and included accessories. Operating conditions are an ambient temperature of 0 to 40 C and relative humidity not greater than 70%.
The tester provides AC/DC output voltage ranges of 10 to 25 kV on the 50 kV range and 50 to 100 kV on the 100 kV range, with a 5 kVA transformer capacity. It supports voltage and current presetting, automatic voltage rise, a 1 to 999 second timer or manual testing, and AC 220 V ±10% 50 Hz power. Precautions include good grounding, insulation for the operator, and discharging the high-voltage capacitor after DC tests. This manual is written in Chinese.
| Operating conditions | Ambient temperature 0 to 40 C; relative humidity not greater than 70% |
|---|---|
| Output voltage range and accuracy | AC/DC: 50 kV range 10-25 kV ±5%; 100 kV range 50-100 kV ±5% |
| Output current and accuracy | AC: 0.5-2 mA and 2-20 mA; DC: 0.5-2 mA and 2-10 mA; ±5% ±3 digits |
| Transformer capacity | 5 kVA |
| Voltage preset range and accuracy | AC/DC: 10-50 kV continuous preset; 50 kV range 10-50 kV ±10%; 100 kV range 50-100 kV ±10% |
| Current preset range and accuracy | AC: 0.5-20 mA ±5%; DC: 0.5-10 mA ±5% |
| Timer range and accuracy | 1 to 999 seconds, ±5%; manual testing available; continuous working time less than 10 minutes |
| Voltage rise method | Automatic |
| Working voltage | AC 220 V ±10%, 50 Hz |
| Voltage rise time | 50 kV range: 0 to 50 kV in about 45 s; 100 kV range: 0 to 100 kV in about 45 s |
| Power socket | 220 V AC, built-in 20 A fuse |
It requires AC 220 V ±10% at 50 Hz.
On the 50 kV range it is 10 to 25 kV, and on the 100 kV range it is 50 to 100 kV, with ±5% accuracy.
The timer range is 1 to 999 seconds with ±5% accuracy, and manual testing is also available.
The high end of the high-voltage capacitor (red clip) must be grounded and discharged.