Truwel Ecological Meteorological Instrument Field Calibration Technical Presentation

Beijing Huayirui Technology Co., Ltd. is a high-tech enterprise established in 2007 with 60 employees. It serves the fields of meteorology, hydrology, environmental protection, scientific research, agriculture, energy, national defense, etc., and provides field calibration of ecological meteorological instruments, observation manuals and observation specifications, calibration cycles, data acquisition. The calibration of the instrument, etc.


Truwel Ecological Meteorological Instrument Field Calibration Technical Presentation - cover page
Brand
Truwel
Document type
Application Note
Language
Chinese
Pages
36
File format
PDF
File size
6.2 MB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
3796877069E4E017AD69585F94B34B34

About this manual

This Chinese presentation, dated 2019-07-25 and published by Beijing Huayirui Technology Co., Ltd. (www.truwel.com), deals with field calibration of ecological meteorological instruments. Its outline lists the verification intervals and requirements found in surface meteorological specifications, several field calibration methods, a mobile calibration and maintenance system, field accuracy verification methods, and calibration of energy balance systems with parallel comparison of gradient systems.

The slides cite Chinese Meteorological Administration and WMO surface observation specifications, the CERN and CFERN long-term observation specifications and LY/T standards, and give typical verification intervals for wind, precipitation, temperature, humidity and pressure sensors. Equipment covered includes the TRU-W starting wind speed calibrator, the TRU-WS wind speed calibrator, the HFP01SC self-calibrating soil heat flux sensor system, rain and snow gauge maintenance, and parallel comparison with HMP155A and WindObserve sensors. This manual is written in Chinese.

What's inside

  • Verification intervals and requirements in surface meteorological specifications
  • Data logger verification
  • Field calibration methods for anemometers
  • TRU-W starting wind speed calibrator
  • TRU-WS wind speed calibrator
  • Rain and snow gauge maintenance
  • HFP01SC self-calibrating soil heat flux sensor system
  • HMP155A parallel comparison
  • WindObserve two-dimensional wind parallel comparison
  • Field accuracy verification and mobile calibration system

Specifications

TRU-W verification range0 m/s to 10.0 m/s
TRU-W wind speed resolution0.04 m/s
TRU-W maximum permissible error±0.05 m/s
TRU-WS operating temperature-20 to 60 °C
TRU-WS wind speed detection range0.3 to 40 m/s
TRU-WS power supplyMains or lithium battery
TRU-WS standby time on batterynot less than 4 h at full load
HFP01SC sensitivity50 µV/W.m-2
HFP01SC temperature range-30 to 70 °C

Frequently asked questions

How often must wind speed and wind direction sensors be verified?

The verification interval for wind speed and wind direction sensors is generally 2 years.

What is the verification interval for humidity sensors?

Humidity sensors are generally verified once a year.

Do fully digital stations need data logger verification?

No, fully digitalized stations do not need verification; stations with analog acquisition do.

What is the verification range of the TRU-W starting wind speed calibrator?

It covers 0 m/s to 10.0 m/s, with a wind speed resolution of 0.04 m/s.