WHS-05 heat sealing tester adopts hot pressing sealing method to measure the parameters of heat sealing temperature, heat sealing time and heat sealing pressure of plastic film substrate, soft packaging composite film, coated paper and other heat sealing composite film, providing users with accurate heat sealing test indicators. The control system is fully digitalized, and the key components are international famous brand products. The automation degree is high and the operation is simple.
This two-page leaflet from Nanbeichao (南北潮商城, www.nbchao.com) introduces the WHS-05 heat sealing tester. The instrument uses a hot-press sealing method to determine heat sealing temperature, time and pressure parameters for plastic film substrates, flexible packaging composite films, coated paper and other heat-sealable composite films. The text states that the control system is fully digital and that key components are sourced from internationally known brands.
The document lists eleven product features, including digital P.I.D. temperature control, aluminum alloy sealing jaws, a twin-cylinder structure with adjustable pressure balance, pneumatic control elements, anti-scald and leakage protection design, automatic and manual working modes, and cooling fans on both sides. A technical parameters table gives the temperature, pressure, time, heating mode, air supply, power, dimensions, weight, standard accessories and the standards the instrument follows. This manual is written in Chinese.
| Heat sealing temperature | Room temperature +10 to 250 °C |
|---|---|
| Temperature control accuracy | ±0.5 °C |
| Heat sealing pressure | 0–0.7 MPa |
| Heat sealing time | 0.01–99.99 s |
| Sealing surface | 40*10*5 块 |
| Heating method | Single or double heating; upper and lower sealing jaws can be switched and controlled separately |
| Test modes | Manual mode / automatic mode |
| Air supply pressure | ≤0.7 MPa (air supply provided by user) |
| Power supply | 220 V 50 Hz |
| Dimensions | 550*330*460 mm |
| Weight | Approx. 35 kg |
| Standard accessories | Main unit, foot switch, air supply line, air filter, power cord, manual |
| Standards | QBT 2358, YBB 00122003, ASTM F2029 |
The temperature control accuracy is stated as ±0.5 °C.
The heat sealing time can be set from 0.01 to 99.99 seconds.
Yes. The air supply pressure should be ≤0.7 MPa and the air source is provided by the user.
The leaflet lists QBT 2358, YBB 00122003 and ASTM F2029.