NRSH-C insulating oil dielectric loss tester is a high-precision, digital industrial instrument developed by Wuhan Nanyue Electric Co., Ltd. in accordance with IEC and national regulations, based on its own technical advantages. The instrument adopts high-frequency induction heating technology, uniform heating, fast heating speed and short time. The single-chip microcomputer is used for control and calculation to realize automation, high accuracy of capacitance and loss measurement, stable data and good repeatability. The output voltage is 0-2000V continuously adjustable, suitable for the different voltage levels of different test samples. It is widely used for the measurement of dielectric loss of transformer oil, capacitor oil and cable oil.
The NRSH-C is an industrial instrument made by Wuhan Nanruo Electric for measuring the dielectric loss of insulating oils, including transformer oil, capacitor oil and cable oil. According to the manual, it uses high-frequency induction heating and microcontroller control and calculation, and its output voltage is continuously adjustable from 0 to 2000 V.
This user manual gives safety measures, performance features, technical specifications and front and rear panel descriptions. It also explains how to assemble, clean and use the oil cup, lists precautions, and covers standard accessories and after-sales service. This manual is written in Chinese.
| Measurement range | Capacitance: 1 pF to 200 pF; dielectric loss: 0 to 10% |
|---|---|
| Capacitance accuracy | ±(0.5%·C + 1 pF) |
| Dielectric loss accuracy | ±(1%·D + 0.00008) |
| Resolution | Capacitance: 0.01 pF; dielectric loss: 1×10-5 |
| Temperature control range | Room temperature to 100°C |
| Temperature control accuracy | ±0.5°C |
| Output voltage | 500 V to 2000 V, continuously adjustable |
| Working power | AC 220 V ±10%, 50 Hz |
| Weight | 10 kg |
| Oil cup electrode gap | 2 mm |
| Oil cup empty capacitance | 60 ±5 pF |
| Oil cup liquid capacity | 40 ml |
It is suitable for dielectric loss measurement of transformer oil, capacitor oil and cable oil.
The capacitance measurement range is 1 pF to 200 pF.
About 130 ml of oil must be injected into the temperature measurement hole.
Chemical-grade petroleum ether with a distillation range of 60 to 90°C is specified for thorough cleaning.