Instruction manual of microcomputer Marshall asphalt mixture stability tester, mainly including product technical parameters, operation instructions, printing, manual rise / fall, calibration and so on
This manual describes the ChenXin MSY-90 microcomputer Marshall asphalt mixture stability tester, including its main technical parameters, operating steps, printing, manual raising and lowering, and calibration procedure.
The document also includes a packing list and a certificate of conformity with inspection items such as maximum load, test space, specimen size, motor power, flow value range, rise rate, and noise. This manual is written in Chinese.
| Pressure measurement range | 0-50 kN |
|---|---|
| Pressure display unit | 0.01 kN (10 N) |
| Flow value measurement range | 0-15.00 mm |
| Working voltage | 220 V, 50 Hz |
| Working temperature | 5-35 °C (recommended 10-30 °C) |
| Maximum load | 50 kN |
| Test space | 300 mm |
| Specimen size | 101.6-152.4 mm |
| Motor power | 370 W |
| Flow value range (certificate) | 0-10 mm |
| Rise rate | 50 ± 5 mm/min |
| Noise | ≤75 dB |
The pressure measurement range is 0-50 kN, displayed in 0.01 kN (10 N).
The working voltage is 220 V at 50 Hz.
The instrument enters overpressure protection, stops the test, and lowers the control part to a certain height.
The pressure calibration password is 0001 and the displacement calibration password is 0002.