Triangular weir user manual, introduces the material, function, installation requirements and precautions of triangular weir
This is the product instruction booklet for the JK-TB02 triangular weir made by Zhejiang Jingkong Instrument Co., Ltd. It covers unpacking checks, storage conditions, safety precautions, the materials and function of the weir, and the requirements for installing it in an open channel. The weir is used together with the GFM-2 ultrasonic open-channel flow meter to convert flow in a channel into a liquid level.
The document gives installation notes such as aligning the weir centerline with the channel centerline, and a table of dimensions (B=600, He=250, P=250 mm) for one weir size with a matching channel size of 600×500. Diagrams show the weir plate shape and the installation arrangement in the channel. The text states that the manual does not cover special technical specifications, repair, or explosion-proof and intrinsically safe operation. This manual is written in Chinese.
| Material | PVC, fiberglass, stainless steel (optional) |
|---|---|
| Storage temperature | (-30 to +60) °C |
| Storage humidity | (5 to 80) %RH, avoid condensation |
| Weir dimensions (table) | B = 600 mm, He = 250 mm, P = 250 mm |
| Matching channel size | 600 × 500 mm |
| Compatible flow meter | GFM-2 ultrasonic open-channel flow meter |
| Maximum flow (printed in table) | 162 per hour (unit as printed) |
It is used together with the GFM-2 ultrasonic open-channel flow meter, which converts the flow in the channel into a liquid level for measurement.
The weir is available in PVC, fiberglass and stainless steel; thicker material is used as flow increases.
Keep it in its original packing box in as-delivered condition, away from rain and places with large temperature or humidity differences, at -30 to +60 °C and 5 to 80 %RH.
The centerline of the weir must coincide with the centerline of the channel, and the attached meter's control unit should be mounted 0.5 to 1 m above the weir plate.