JH-STA350 is a synchronous thermal analysis instrument that can measure the melting, crystallization, phase transition, reaction temperature and reaction heat, burning heat, specific heat and other characteristics of materials.
The JH-STA350 is a synchronous thermal analyzer that combines thermogravimetric analysis (TG) with differential thermal analysis (DTA) or differential scanning calorimetry (DSC) in one measurement on the same sample. It measures DSC characteristics such as melting, crystallization, phase transitions, reaction temperature and heat, combustion heat, and specific heat, and TG characteristics such as thermal stability, decomposition, oxidation/reduction, adsorption/desorption, free and crystalline water content, and composition ratios.
Its temperature range is room temperature to 1350°C, with 0.1°C resolution and ±0.1°C fluctuation. The heating rate is 1 to 80°C/min, cooling from 1000°C to 100°C takes 15 minutes, and the balance measures 1 mg to 2 g (expandable to 30 g) with 1 μg resolution. The unit has a 24-bit color 7-inch LCD touch screen, supports inert, oxidizing, reducing, static, and dynamic atmospheres, and uses a standard USB interface. This manual is written in Chinese.
| Temperature range | Room temperature to 1350°C |
|---|---|
| Temperature resolution | 0.1°C |
| Temperature fluctuation | ±0.1°C |
| Heating rate | 1 to 80°C/min |
| Temperature control modes | Heating, constant temperature, cooling |
| Cooling time | 15 min (1000°C to 100°C) |
| Balance measurement range | 1 mg to 2 g, expandable to 30 g |
| Constant-temperature time | 0 to 300 min, arbitrarily set |
| Display | 24-bit color, 7-inch LCD touch screen |
| Atmosphere | Inert, oxidizing, reducing, static, dynamic |
| Atmosphere device | Built-in gas flow meter with two-gas switching and flow control |
| Software | Smart software automatically records TG curves, processes data, and prints test reports |
| Data interface | Standard USB interface |
| Power | AC 220 V, 50 Hz |
| Sensor | Built-in high-temperature platinum sensor |
The temperature range is room temperature to 1350°C.
The heating rate is 1 to 80°C/min.
It supports inert, oxidizing, reducing, static, and dynamic atmospheres.
The balance measures 1 mg to 2 g and can be expanded to 30 g.