Ligent L600 Multifunctional Intelligent Transmitter Instruction Manual

L600 multifunctional intelligent transmitter is a high-performance, high-precision fast data acquisition product, which also has peak capture, comparative output, analog output and communication output functions. The module is small in size and easy to install. It can be used with our dedicated sensors to adapt to various production systems for force measurement and precision control.


Ligent L600 Multifunctional Intelligent Transmitter Instruction Manual - cover page
Brand
Ligent
Category
pressure transmitters
Document type
User Manual
Language
Chinese
Pages
11
File format
PDF
File size
1 MB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
F14B47E0E36001A8527CDFFD82B1C65F

About this manual

The L600 from Shenzhen Lizhun Sensor Technology Co., Ltd. (website www.ligentcn.com) is a multifunctional intelligent transmitter for fast force data acquisition, with peak capture, comparison output, analog output and communication output. Used together with the company's dedicated sensors, it is intended for force measurement and precision control in production systems.

This Chinese instruction manual (version V1.0) covers the main functions, technical specifications, product model, dimensions and terminal definitions, function descriptions, the Modbus-RTU communication protocol, the parameter table, the communication address table and the digital I/O, followed by safety notes, a statement and warranty information. This manual is written in Chinese.

What's inside

  • Overview
  • Main Functions
  • Technical Specifications
  • Product Model
  • Dimensions and Terminal Definitions
  • Function Description
  • Communication Protocol
  • Parameter Table
  • Communication Address Table
  • Digital Input/Output

Specifications

Acquisition performance24-bit acquisition chip, 5–3200 Hz acquisition speed; effective accuracy 17–18 bits at 240 samples per second
Sensor interface5 V, 100 mA external supply; 5-wire sensor with TEDS (compatible with ordinary sensors)
Communication portRS485, 1200–115200 bps, Modbus-RTU or active sending
Digital I/O1 DI + 3 DO, optically isolated, OC output 200 mA/24 V
Analog output4–20 mA (below 500) or 0–10 V / ±5 V
Power supplyDC 24 V ±10%, < 200 mA (separate 24 V DC supply required)
Operating environmentOperating -10 to 40 °C, storage -20 to 60 °C, humidity below 85 %RH

Frequently asked questions

Which communication protocols does the L600 support?

The external RS485 port can run the Modbus-RTU protocol and a custom protocol; the internal TTL port for auxiliary setup is fixed to Modbus-RTU.

What analog output options are available?

The analog output can be selected as 4–20 mA or 0–10 V / ±5 V, and its relationship to the measured value can be set.

How does TEDS calibration work?

When the TEDS enable parameter is valid, the module reads the TEDS data automatically at power-up and completes calibration; if not, the previous coefficients are used.

How can the transmitter be restored to factory settings without communication?

Remove the two countersunk screws from the bottom, open the housing, short the first three holes of the JPP0 header and power up; the module returns to mode 1.

Download PDF (1 MB)

More Ligent Manuals


View all Ligent manuals →