L600 multifunctional intelligent transmitter is a high-performance, high-precision fast data acquisition product, which also has peak capture, comparative output, analog output and communication output functions. The module is small in size and easy to install. It can be used with our dedicated sensors to adapt to various production systems for force measurement and precision control.
The L600 from Shenzhen Lizhun Sensor Technology Co., Ltd. (website www.ligentcn.com) is a multifunctional intelligent transmitter for fast force data acquisition, with peak capture, comparison output, analog output and communication output. Used together with the company's dedicated sensors, it is intended for force measurement and precision control in production systems.
This Chinese instruction manual (version V1.0) covers the main functions, technical specifications, product model, dimensions and terminal definitions, function descriptions, the Modbus-RTU communication protocol, the parameter table, the communication address table and the digital I/O, followed by safety notes, a statement and warranty information. This manual is written in Chinese.
| Acquisition performance | 24-bit acquisition chip, 5–3200 Hz acquisition speed; effective accuracy 17–18 bits at 240 samples per second |
|---|---|
| Sensor interface | 5 V, 100 mA external supply; 5-wire sensor with TEDS (compatible with ordinary sensors) |
| Communication port | RS485, 1200–115200 bps, Modbus-RTU or active sending |
| Digital I/O | 1 DI + 3 DO, optically isolated, OC output 200 mA/24 V |
| Analog output | 4–20 mA (below 500) or 0–10 V / ±5 V |
| Power supply | DC 24 V ±10%, < 200 mA (separate 24 V DC supply required) |
| Operating environment | Operating -10 to 40 °C, storage -20 to 60 °C, humidity below 85 %RH |
The external RS485 port can run the Modbus-RTU protocol and a custom protocol; the internal TTL port for auxiliary setup is fixed to Modbus-RTU.
The analog output can be selected as 4–20 mA or 0–10 V / ±5 V, and its relationship to the measured value can be set.
When the TEDS enable parameter is valid, the module reads the TEDS data automatically at power-up and completes calibration; if not, the previous coefficients are used.
Remove the two countersunk screws from the bottom, open the housing, short the first three holes of the JPP0 header and power up; the module returns to mode 1.