The Photoreactor m2 is an integrated benchtop instrument that accelerates photochemical redox catalysis reactions. It has a modular design that allows the application of wavelengths from 365nm to 450nm and maximizes photon capture area through a 360-degree reflective environment.
The Photoreactor m2 is a benchtop instrument for chemists and researchers that uses photoredox catalysis to accelerate chemical reactions. It combines LED light, mechanical stirring, and cooling; users can set temperature, light intensity, fan speed, stirring speed, and time.
The manual covers safety guidelines, system overview, installation, operation, system information, care and handling, maintenance, technical data, and product classification. The m2 has a light shield with an interlock, a touchscreen interface, interchangeable LED modules for 365 nm to 450 nm, and K-type thermocouple temperature feedback; sample vial sizes are gc, 4, 8, 20, and 40 ml. This manual is written in Chinese.
| Width, cover closed | 12.2 cm (4.8 in) |
|---|---|
| Height, cover closed | 27.2 cm (10.7 in) |
| Depth, cover closed | 28.2 cm (11.1 in) |
| Height, cover open | 33.8 cm (13.3 in) |
| Depth, cover open | 42.67 cm (16.8 in) |
| Weight | Approx. 2 kg (4.4 lb) |
| Operating mode | Continuous |
| Mains cable | 10 A/250 V |
| Power supply | 85-264 V, 50/60 Hz, 120 VA |
| Expected service life | 5 years |
| Operating temperature | 5 to 40 °C (41 to 104 °F) |
| Operating air pressure | 700 hPa to 1060 hPa |
The default LED intensity is 50%.
The supported sample vial sizes are gc, 4, 8, 20, and 40 ml.
The LED stops emitting and the fan and stirring magnet stop rotating.
It must use the supplied power supply rated 85-264 V, 50/60 Hz, full range.