Y-Tec red blood cell lysis liquid is a red blood cell lysis liquid with optimized formula, which is used to lyse red blood cells and almost does not damage lymphocytes or other cells with nuclei.
The Y-Tec Red Blood Cell Lysis Buffer is supplied as catalog number YLC0500 in a 500 ml size by Shanghai Qichun Biotechnology Co., Ltd. The instructions state that the optimized formulation lyses red blood cells while causing almost no damage to lymphocytes or other nucleated cells. Its main active ingredient is ammonium chloride, and it is not suitable for lysing nucleated red blood cells such as bird or poultry red blood cells.
The manual gives procedures for tissue cell samples and blood samples, each with a standard washing method and a rapid method without washing. Steps include adding 3–5 or 6–10 times the cell volume of lysis buffer, lysing for 1–2 minutes, centrifuging at 400–500g for 5 minutes, and washing with PBS, HBSS, saline, or serum-free culture medium. Notes cover longer lysis times for human peripheral blood and the option to skip the first centrifugation for small blood samples. This manual is written in Chinese.
| Product name | Y-Tec Red Blood Cell Lysis Buffer |
|---|---|
| Catalog number | YLC0500 |
| Size | 500 ml |
| Main active ingredient | Ammonium chloride |
| Not suitable for | Nucleated red blood cells, such as bird or poultry red blood cells |
The main active ingredient is ammonium chloride.
No. The instructions state that it is not suitable for lysing nucleated red blood cells, such as bird or poultry red blood cells.
Centrifuge at 400–500g for 5 minutes and discard the red supernatant.
For mouse blood, 1–2 minutes is sufficient; human peripheral blood may be lysed for 4–5 minutes.