About this manual

The AnGui AN-100A Oxygen Index Tester is an instrument designed to measure the oxygen concentration (volume percentage) required to sustain the combustion of polymers. It complies with the national standard GB2406 and is suitable for testing the combustion performance of materials such as plastics, rubber, and foams. The device determines the minimum oxygen volume percentage in an oxygen-nitrogen mixture needed to keep a sample burning for 50mm or for 3 minutes.

The manual covers technical specifications, installation procedures, and usage instructions. Key specifications include a combustion tube inner diameter of 100mm, a height of 450mm, and a working pressure range of 0.05-0.15MPa. The instrument requires an oxygen and nitrogen source with purity not lower than 98% and operates on a 220V AC, 50Hz power supply. Users are instructed to calibrate the oxygen concentration before each test and to follow specific steps for igniting and recording the results. This manual is written in Chinese.

What's inside

  • Overview
  • Technical Parameters
  • Installation
  • Usage
  • Precautions
  • Packing List
  • Certificate

Specifications

Combustion tube inner diameter100mm
Combustion tube height450mm
Flowmeter accuracy2.5 grade
Pressure gauge accuracy2.5 grade
Pressure gauge resolution0.01MPa
Oxygen concentration indicator resolution0.1%
Operating humidity45-75%
Power supply220V, AC, 50Hz

Frequently asked questions

What standard does this instrument comply with?

The instrument complies with the national standard GB2406.

What materials is this tester suitable for?

It is suitable for testing plastics, rubber, and foams.

What is the required purity for the oxygen and nitrogen gas sources?

The purity of the oxygen and nitrogen must not be lower than 98%.

What is the working pressure range of the instrument?

The working pressure range is 0.05-0.15MPa.