AN-100A Oxygen Index Tester is a device used to measure the minimum oxygen content (volume percentage) of oxygen-nitrogen mixed gas required to keep a substance burning for 50 mm or 3 minutes after ignition.
The AnGui AN-100A Oxygen Index Tester is an instrument designed to measure the oxygen concentration (volume percentage) required to sustain the combustion of polymers. It complies with the national standard GB2406 and is suitable for testing the combustion performance of materials such as plastics, rubber, and foams. The device determines the minimum oxygen volume percentage in an oxygen-nitrogen mixture needed to keep a sample burning for 50mm or for 3 minutes.
The manual covers technical specifications, installation procedures, and usage instructions. Key specifications include a combustion tube inner diameter of 100mm, a height of 450mm, and a working pressure range of 0.05-0.15MPa. The instrument requires an oxygen and nitrogen source with purity not lower than 98% and operates on a 220V AC, 50Hz power supply. Users are instructed to calibrate the oxygen concentration before each test and to follow specific steps for igniting and recording the results. This manual is written in Chinese.
| Combustion tube inner diameter | 100mm |
|---|---|
| Combustion tube height | 450mm |
| Flowmeter accuracy | 2.5 grade |
| Pressure gauge accuracy | 2.5 grade |
| Pressure gauge resolution | 0.01MPa |
| Oxygen concentration indicator resolution | 0.1% |
| Operating humidity | 45-75% |
| Power supply | 220V, AC, 50Hz |
The instrument complies with the national standard GB2406.
It is suitable for testing plastics, rubber, and foams.
The purity of the oxygen and nitrogen must not be lower than 98%.
The working pressure range is 0.05-0.15MPa.