This manual mainly introduces the features, installation, storage, technical indicators, structure, function, operation and maintenance guide of KPA600 series high precision power analyzers of Kefuna Technology.
This user manual is for the Kefuna KPA600 series high precision power analyzer. It describes safety regulations, installation and transport conditions, technical specifications, panel and interface details, measurement functions, operating procedures, and maintenance guidance.
The specifications in the manual list KPA610B, KPA610, KPA610A, and KPA600 models, with a 600 V AC voltage range, 20 A AC current range, and 100 kHz frequency range. This manual is written in Chinese.
| Input voltage | 220 V AC ±10%, 50/60 Hz |
|---|---|
| Input fuse | 0.5 A |
| Voltage range | 600 V AC |
| Current range | 20 A AC |
| Frequency range | 100 kHz |
| Current resolution | 0.0001 mA at 5 mA rated range |
| Voltage rating | 600 V |
| Current rating | 20 A |
| Voltage measurement ranges | 15/30/60/150/300/600 V |
| Sampling frequency | about 100 kS/s |
| Basic measurement accuracy | 0.1 or 0.2 class depending on model |
| Active power measurement range | 75 mW to 12 kW |
| Active power resolution | 0.1 mW |
| Frequency measurement range | 0.5 Hz to 100 kHz or 0.5 Hz to 10 kHz depending on model |
| Frequency measurement accuracy | ±0.1 of displayed value |
| Harmonic measurement | 10 Hz to 1.2 kHz; 1st to 50th harmonic content and total distortion |
| Energy measurement range | 0 to 9999 MWh |
| Energy measurement accuracy | ±0.5% of displayed value |
| Bandwidth | voltage bandwidth less than 100 kHz or less than 10 kHz depending on model |
| Dimensions | 361 x 213 x 99 mm |
| Working environment | 20 ±10 °C, relative humidity below 75% |
| Storage environment | -40 to 55 °C, 20 to 80% RH at 40 °C |
| Measured quantities | voltage, voltage frequency, current, active/reactive/apparent power, power factor, energy, harmonics |
| Data storage | USB storage and parameter setting storage/copy |
| Communication interfaces | RS232; PLC remote interface and RS485 optional |
The analyzer uses a single-phase 220 V AC ±10%, 50/60 Hz power supply.
The input fuse is 0.5 A.
The specification table lists KPA610B, KPA610, KPA610A, and KPA600.
It is intended for household appliance, lighting, motor, and transformer manufacturers, as well as laboratories, new product development, and quality supervision departments.
It must be operated by trained, qualified personnel.