Standard desktop laser measurement instrument, which can be widely used in cable, enameled wire, optical fiber, micro wire drawing, various pipes and rods, machining and other industries.
This is the Chinese instruction manual (说明书) for the Kecheng standard desktop laser diameter gauge (标准台式激光测径仪) from Ningbo Beilun Kecheng Instrument Co., Ltd. The document covers the LDM-01A, LDM-01B, LDM-03A, LDM-03B, LDM-10A, LDM-10B, LDM-30A, LDM-30B, LDM-60A, LDM-60B, LDM-90A and LDM-90B models, each made up of a sensor and a display control unit in a split two-part construction. The text states it can be used for cables, enameled wire, optical fiber, fine drawing, various tubes and rods, and machining work.
The manual gives the measuring ranges, accuracy, repeatability and resolution table for all models, then describes the appearance and dimensions, the panel keys and indicator lamps, and the rear-panel output terminals. It closes with installation notes and step-by-step operating instructions for switching on, setting the nominal value, setting the upper and lower deviation limits, calibrating the machine with standard rods, the data hold function and the out-of-tolerance alarm. This manual is written in Chinese.
| Measuring range (LDM-01A) | 0.05000–1 mm |
|---|---|
| Accuracy and repeatability (LDM-01A) | ±0.5 µm accuracy, ±0.2 µm repeatability, 0.00001 mm resolution |
| Measuring range (LDM-90B) | 0.5000–90 mm |
| Accuracy and repeatability (LDM-90B) | ±5 µm accuracy, ±3 µm repeatability, 0.0001 mm resolution |
| Scanning speed | 1800 measurements per second |
| Data processor | High-performance DSP processor for high-speed, high-accuracy calculation |
| Communication interface | RS232 or RS485, can be connected to a computer for online measurement |
| Sensor dimensions (small desktop models) | 160 mm × 70 mm × 90 mm |
| Sensor dimensions (standard models) | 448 mm × 120 mm × 75 mm |
| Display unit dimensions | 130 mm × 140 mm × 150 mm |
The manual states a scanning speed of 1800 measurements per second.
All models first enter a waiting state and the screen shows "123456" for about one second before entering normal measurement.
It has an RS232 or RS485 communication interface that can be connected to a computer for online measurement and data recording.
Enter calibration mode and place the 2 mm standard rod at the center of the measuring area, then after the data stabilizes place the 8 mm rod in the V-groove or guide wheel center, pressing the ▲ key each time.