3168G type electric transformer load tap changer parameter comprehensive testing instrument is a special instrument used to measure and analyze the dynamic parameters of transformer load tap changer. It can realize the test of load tap changer's transition process, transition waveform, transition time, each instantaneous transition resistance value and three phase synchronization.
This Chinese user manual covers the Jingyi 3168G, a tester for measuring and analyzing the operating characteristics of a power transformer's on-load tap changer. According to the text it records the transition process, transition waveform, transition time, the transition resistance at each instant and three-phase timing, and it can be used with or without the transformer winding connected. It is an improved version of the company's earlier 3168C, made by Baoding Jingyi Electronic Instrument Co., Ltd.
The manual gives the use conditions and safety notes, the main technical data and the front panel layout, then walks through wiring the three phase leads and the neutral lead, setting the tap range, triggering the test and reading the waveform with the cursor lines. It also describes clock adjustment, printing the waveform and the statistical average resistance with the thermal printer, storing results (three sets in internal memory plus a USB drive option) and screen paging for waveforms longer than 60 ms. The last pages list MR company type designations and the company address. This manual is written in Chinese.
| Transition time test range | 0 – 240 ms |
|---|---|
| Transition resistance range, low current | 400 mA, 0.4 – 40 Ω |
| Transition resistance range, high current | 1 A, 0.3 – 20 Ω |
| Transition resistance accuracy | ±(5 % + 0.1 Ω) |
| Environment temperature / humidity | -20 °C to +40 °C, humidity ≤ 80 % |
| Supply voltage | 220 V |
| Internal memory | 32 K |
| Internal result storage | 3 sets, retained after power off |
| PC software operating systems | WIN98, WIN2000 or XP |
Yes. The tester can be used with or without the winding connected, but testing on a live circuit is strictly forbidden.
The transition time test range is 0 to 240 ms.
Results can be saved, printed with the built-in printer, or stored to a USB drive; the instrument itself holds three sets that are kept after power off.
It is connected to a 220 V supply together with the test leads.