LXH hot (cold) quantity flow meter is a kind of flow, cold and hot quantity flow meter that can be equipped with various field flow sensors and displayed by 12864 graphic dot matrix LCD screen. It can measure the mass flow rate, inlet and outlet temperature, pressure, heat and cold quantity of hot water, cold water, saturated steam, oil and other media. It can realize pressure, temperature automatic compensation density, temperature automatic compensation enthalpy, and has day report, month report, coming power failure record, power failure memory, password setting, protocol measurement and communication networking functions. If an external miniature printer is attached, it can print cumulative data at a fixed time.


Shanghai Jiahu Instrument LXH Heat (Cold) Quantity Flow Meter Operation Instructions - cover page
Brand
Shanghai Jiahu Instrument
Document type
User Manual
Language
Chinese
Pages
28
File format
PDF
File size
460 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
2C0224A32A0A414D955A5F6F3295EA8D

About this manual

This document provides operation instructions for the LXH Heat (Cold) Quantity Flow Meter by Shanghai Jiahu Instrument. It covers installation, wiring diagrams, display functions, user programming, communication protocols, and calibration procedures. The meter measures mass flow, temperature, pressure, and heat/cold quantities with automatic compensation and data logging capabilities. It features a 12864 LCD display, RS485/RS232 communication, and optional printer support for data output. This manual is written in Chinese.

What's inside

  • LXH Instrument Introduction
  • Main Performance Indicators
  • Unpacking and Contents
  • Installation
  • Wiring Diagram
  • Display Functions
  • User Programming
  • Communication Functions
  • Programming Examples
  • Accuracy Calibration

Specifications

Flow Input Signal0~10 mA, 4~20mA, or 0~5000Hz
Pressure Input Signal0~10 mA, 4~20mA
Temperature Input SignalPt1000
Output Signal4~20mA analog or 50V/30mA switch
Measurement AccuracyBetter than 0.2% for analog, 0.1% for frequency
Working PowerAC 220V±15%, <5W
Dimensions160×80×150 mm (W×H×D)
WeightApprox. 1.0kg
Operating Temperature5~50℃

Frequently asked questions

What communication interfaces does the LXH meter support?

It supports RS485 or RS232 communication interfaces with lightning protection.

What is the working power requirement for the LXH meter?

It requires AC 220V±15% with a power consumption of less than 5W.

What is the display capability of the LXH meter?

It displays cumulative and instantaneous flow, heat/cold quantities, pressure, density, temperature, and time.