The HTJ-16 Automatic Backlash Angle Tester Product Operation Manual of Changsha Xinxiong Instrument Technology Co., Ltd., this instrument is mainly used to test the backlash of conductors with a nominal diameter of 0.080 mm or above 1.600 mm and below lacquered round wires. The high-end single-chip microcomputer is controlled, the running speed is fast, and the anti-interference ability is strong. The winding 5 laps stops automatically. Various size samples do not need to adjust the sensor. When the backlash, the coil slowly loosens, without sudden backlash or sudden jump, the LED automatically records and displays the test results, and the display value and the scale plate indicate the error is not more than ±1°. The operation is simple, the sensitivity is high, and the repeatability is good.
The Xin Xiong HTJ-16 Automatic Backlash Angle Tester is designed to detect the resilience of enameled round wires with a nominal diameter from 0.080mm to 1.600mm. The instrument operates under high-end microcontroller control, featuring automatic winding of 5.5 turns and a slow, controlled unwinding process to prevent sudden jumps. It includes an LED display for recording results with an accuracy error of no more than ±1° compared to the dial indicator.
The manual details the setup and operation steps, including power connection, resetting the scale, selecting appropriate winding rods and weights based on wire diameter, and adjusting the light-blocking device. It also outlines the testing procedure: winding the sample, positioning the tail end to trigger the sensor, and initiating the automatic return to measure the backlash angle. Safety precautions and a list of included accessories are also provided. This manual is written in Chinese.
| Test Standard | GB/T4074.3-2008/IEC60851-3 |
|---|---|
| Inspection Standard | JB/T4279.4-2008 |
| Applicable Wire Diameter | 0.080mm to 1.600mm |
| Winding Speed | 7±0.5 r/min |
| Input Power | AC220V±10% 50Hz |
| Power Consumption | <80W |
| Dimensions | 560×314×360mm |
| Weight | 25kg |
| Display Error | ±1° |
The instrument is used to test enameled round wires with a nominal diameter of 0.080mm or larger and 1.600mm or smaller.
The instrument automatically stops after winding 5.5 turns.
The error between the displayed value and the dial indicator indication does not exceed ±1°.