Industrial ion controller is a high intelligent, multi-function online chemical analyzer, which can accurately measure the ion value and temperature of the solution. It is suitable for the accurate measurement of the ion content in the high-pressure steam boiler feedwater, mineral water, drinking water, surface water and seawater of power plants and steam power plants.
This manual provides detailed instructions for operating and configuring the PR-300B-LXX-N01-* industrial ion controller. It includes sections on device installation, wiring, menu navigation, alarm settings, analog output calibration, and ModBus communication protocols. The controller measures ion concentrations and temperature with configurable parameters and supports RS485 communication for integration into monitoring systems.
The device features a 128x64 LCD display, 4-20mA current output, and built-in data storage for historical measurements. It is designed for applications such as boiler feedwater, drinking water, and seawater analysis, with configurable ion-specific measurement ranges and calibration points. This manual is written in Chinese.
| Ion Measurement Range | 0-100.00mg/L (resolution 0.01mg/L, no support for chloride), 0-1000.0mg/L (resolution 0.1mg/L, supports Na, Mg, Ca), 0-3500.0mg/L (resolution 0.1mg/L, supports chloride) |
|---|---|
| Temperature Measurement Range | -20.0-80.0℃ |
| Temperature Measurement Error | ±0.2℃ |
| Input Impedance | ≥1012Ω |
| Output Current Error | ≤±1%FS |
| Data Storage | 72 points, continuous 3 days data |
| RS485 Communication | ModBus-RTU protocol |
| Power Supply | DC24V/AC220V (optional) |
| Power Consumption | DC24V ≤1.6W; AC220V ≤15W |
| Protection Rating | IP54 |
| Dimensions | 96×96×110 mm |
| Mounting Hole Size | 92×92 mm |
| Weight | ≤0.6kg |
| Operating Temperature | -20~60℃ |
| Relative Humidity | <85%RH |
The chloride ion measurement range is 0-3500.0mg/L with a resolution of 0.1mg/L.
The default ModBus address is 0x01.
The device can store up to 72 data points over three days.
The output current error is ≤±1%FS.