This document introduces the nine-point testing method of the Lianxin brand needle detector, including the introduction of the detector's probe structure, testing frame and test card, and the testing steps of the detector.
This manual details the nine-point calibration procedure for Lianxin JZQ-630K tunnel-type needle detectors using standard test cards with iron balls of varying diameters. It explains how to test sensitivity at nine positions (A1–A9) using a calibration rack and test cards, with specific steps for each point. The manual also notes that sensitivity varies by machine model and includes instructions for different modes (B and C) and troubleshooting for magnet coil issues.
The test kit includes four test cards (Fe0.8mm, Fe1.0mm, Fe1.2mm, Fe1.5mm) and a calibration rack with two height options (9cm and 12cm). Calibration should be performed every three hours to ensure accurate detection performance. This manual is written in Chinese.
| Test Card Diameter | Fe0.8mm, Fe1.0mm, Fe1.2mm, Fe1.5mm |
|---|---|
| Calibration Rack Height | 9cm (for 10cm~15cm) or 12cm (for 15cm~20cm) |
| Test Point Positions | A1 to A9 |
The calibration uses test cards with Fe0.8mm, Fe1.0mm, Fe1.2mm, and Fe1.5mm iron balls.
Calibration should be performed at least every three hours to ensure optimal performance.