HK-228 type hydrazine analyzer instruction manual introduces the function characteristics, operation method and use precautions of the instrument
This manual provides step-by-step instructions for operating and calibrating the HK-228 hydrazine analyzer. It includes details on reagent preparation, programming, and measurement procedures. The device uses photometric colorimetry based on the reaction between hydrazine and p-dimethylaminobenzaldehyde to determine hydrazine concentration. Technical specifications cover measurement range, display, accuracy, and environmental conditions. This manual is written in Chinese.
| Measurement Range | 0~100 µg/L |
|---|---|
| Display | Chinese, dot matrix LCD |
| Instrument Error | ±2% F.S. |
| Resolution | 0.1 µg/L |
| Repeatability | Not more than 1% |
| Stability | ±1% F.S./4h |
| Operating Temperature | 5~45°C |
| Operating Humidity | Not more than 90% RH (no condensation) |
| Dimensions | 420mm × 190mm × 280mm |
| Power Supply | AC 220±22V, 50±1Hz |
The measurement range is 0 to 100 µg/L.
It uses photometric colorimetry based on the reaction between hydrazine and p-dimethylaminobenzaldehyde, following Lambert-Beer’s law.
The instrument has an error of ±2% of full scale.
Calibration involves blank calibration and curve calibration using standard solutions, with stored results for verification.