AO-08 oxygen sensor is an electrochemical sensor specially used for medical devices to measure oxygen concentration, which has the characteristics of fast response and long service life.
The AO-08 oxygen sensor is an electrochemical device designed for medical applications, offering full-range linear output and temperature compensation. It uses a 3.5mm headphone jack interface and requires no external power. The sensor is calibrated and compensated at factory, suitable for devices like oxygen concentrators, ventilators, and incubators. It has a 13-month warranty and operates within 0–50°C, with a lifespan of up to 1.5 million hours at 20°C.
Installation requires manual tightening to ensure airtightness, and the sensor must be kept away from organic solvents. Cleaning is done with distilled water only. The recommended circuit uses ADC readings to calculate oxygen concentration via a formula. This manual is written in Chinese.
| Output Voltage | 9~13 mV (in air) |
|---|---|
| Measurement Range | 0~100 % |
| Response Time (T90) | <15 s |
| Linearity | Full range linear |
| Interface | 3.5 mm headphone jack |
| Operating Temperature | 0~50 ℃ |
| Weight | Approx. 40g |
| Warranty | 13 months |
| Lifespan at 20°C | 1.5 × 10⁶ hours |
The output voltage is 9~13 mV in air.
The response time (T90) is less than 15 seconds.
Clean the sensor with distilled water and let it air dry. Do not use steam sterilization or expose it to chemicals like ethylene oxide or hydrogen peroxide.
The warranty is 13 months from the date of shipment.