Telstar CRYODOS Laboratory Freeze Dryer User's Manual

This manual provides information on how to use and maintain the laboratory freeze dryer.


Telstar CRYODOS Laboratory Freeze Dryer User's Manual - cover page
Brand
Telstar
Category
Pharmaceutical Equipment
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
293 KB
Published
27 October, 2023
Updated
27 October, 2023
MD5 checksum
14969652503C3584CC51A7A9A95EDE80

About this manual

The Telstar CRYODOS is a compact desktop laboratory freeze dryer available in two models: CRYODOS -50 and CRYODOS -80. The manual provides guidelines for handling, operation, and maintenance of the unit, which includes a metal cabinet, vacuum pump, cooling system, and a stainless steel condenser cylinder with eight cocks for vial connections.

The manual details technical specifications such as power requirements (230 V single-phase, 50/60 Hz), condenser temperatures (-50°C or -80°C), ice capacity (4 kg), and refrigeration system (air-cooled compressors). It also covers transport, positioning, electrical connection, vacuum pump assembly, lubrication, and alarm handling.

What's inside

  • General Details
  • Details of the Freeze Dryer
  • Transport
  • Cautions Before Start-Up
  • Working Principle
  • Accessories
  • Operating Instructions
  • Attached

Specifications

Vacuum pump modelVARIAN DS 102
Vacuum pump dimensions430 x 132 x 212 mm
Vacuum pump weight22 kg
Suction flow (50/60 Hz)5/6 m³/h
Motor velocity (50/60 Hz)1500/1800 rpm
Electric consumption (50/60 Hz)0.45/0.55 kW
Oil load capacity0.5 litres
Oil typeAV-30
Ice condenser condensing surface0.136 m²
Ice condenser final temperature (CRYODOS -50)-50°C
Ice capacity (CRYODOS -50)4 kg (1.5 kg/24 h)
Condenser unit materialAISI 316L
Number of compressors (CRYODOS -50)1
Compressor typeSingle-stage hermetic
Refrigerating power (CRYODOS -50)450 W
Refrigerant (CRYODOS -50)R-404a
Assembly refrigerationBy air circulation
Electrical connection230/1/50 and 230/1/60 V/ph/Hz
Power capacity installed (CRYODOS -50)1 kW
Vacuum at pump inlet5 x 10⁻³ mbar
Process vacuum (CRYODOS -50)4 x 10⁻² mbar at -45°C
Vacuum gaugeDigital
Weight (CRYODOS -50)60 kg

Frequently asked questions

What are the power supply requirements for the CRYODOS freeze dryer?

The CRYODOS requires 230 V single-phase + earth at 50 Hz or 60 Hz.

What is the maximum working room temperature?

The maximum working room temperature is +30°C.

How should the vacuum pump oil be handled during long storage?

The vacuum pump oil must be changed to ensure no corrosive waste damages the pump.