SPECTRA 5011/5041 User's Manual
The document serves as a user's manual for the SPECTRA 5011 and SPECTRA 5041 portable alpha-beta-gamma spectrometers, also referred to as the "Surface Investigator" and "Surface Investigator PRO". It covers unpacking, general safety instructions, and maintenance procedures for the device and its components.
The manual explains the principles of alpha and gamma spectroscopy, including geometrical and electronic effects such as dead time and pile-up. It details the theory of operation, configuration parameters like ADC resolution and peak detector delay time, and provides step-by-step instructions for connecting detectors, setting energy ranges, and performing calibration using the included radium dial.
| Power Supply | 220 V (110 V on request) |
|---|---|
| ADC Resolution | 8 to 12 bit |
| CAN Bus Range | up to 1000 meter |
| Energy Range | >30 kev up to about some MeV |
The packing list includes the portable spectrometer, NaJ Scintillation probe, Si-alpha detector or amplifier, CAN Dongle, cables, software on CD, and user manuals.
Clean the surfaces of the SPECTRA and the NaJ scintillation crystal with a dry towel, but do not clean or touch the Si-detector.
Four different delay times are selectable: 7.5µs, 15µs, 30µs, and 60µs.