This manual introduces the IET Labs HACS-Z-A-6E-1pF Decade Capacitance System. The manual includes information on introduction, specifications, operation and maintenance.
This service manual provides detailed instructions for operating, maintaining, and calibrating the HACS-Z-A-6E-1pF Decade Capacitance System. It covers initial setup, switch settings, terminal connections, preventive maintenance, calibration intervals, and adjustment procedures. The system is designed for precise, stable capacitance values from 1 pF to 1.11111 µF, with double-shielded construction for low residual capacitance. It includes specifications, diagrams, and safety guidelines for accurate use and long-term stability.
| Range | 0 to 1.111 110 µF, in 1 pF steps |
|---|---|
| Zero Capacitance | ≤0.1 pF maximum capacitance obtained with all dials set to zero |
| Insulation Resistance | >50,000 MΩ |
| Operating Temperature Range | 10°C to 40°C |
| Dimensions (Bench) | 43.2 cm W x 13.3 cm H x 27.7 cm D (17” x 5.2” x 10.9”) |
| Weight (Bench) | 8.6 kg (19 lb) |
| Accuracy (1 pF-10 µF) | ± (0.05% + 0.5 pF) at 1 kHz |
The range is 0 to 1.111 110 µF, in 1 pF steps.
The maximum zero capacitance is ≤0.1 pF.
Connect the inner shield and outer shield together at the measuring instrument to use it as a 3-terminal capacitance substituter with very low zero capacitance.