This manual describes the installation, operation and troubleshooting of the Hitachi L6200/L6210 Control Module for controlling the Hitachi L6200/L6210 Liquid Chromatograph
This manual describes the DataApex Clarity control module for Hitachi L6200 and L6210 pumps. The module enables direct control of the pump over a serial line from the Clarity environment, including the gradient program, and the instrument method is saved in the measured chromatograms.
It covers requirements, hardware wiring, communication settings, Clarity configuration, method setup, device monitoring, report setup and troubleshooting. Requirements include Clarity with the LC Control module (p/n A24), a free serial COM port and a serial straight DB9F-DB25M cable (p/n SKO3); the pump is controlled by RS232 communication with DIP Switch S1 on its Computer Board.
| Communication interface | RS232 |
|---|---|
| Baud rate | 4800 |
| Data bits and stop bits | 8 databits + 2 stop bits |
| Required cable | Serial straight DB9F-DB25M cable (p/n SKO3) |
| DIP switch location | S1 on Computer Board |
| Required Clarity module | LC Control module (p/n A24) |
It enables direct control of the Hitachi L6200 and L6210 pumps from the Clarity environment over a serial line, including the gradient program.
The DIP switch settings include 8 databits + 2 stop bits and Baud rate 4800; DIP 6 is set OFF for RS232 mode.
The requirements list a serial straight DB9F-DB25M cable (p/n SKO3); the wiring section describes serial cross DB9F-DB25M wiring.
Add or amend the COMMDRV.INI file in the Clarity installation directory to enable communication recording; the created TXT files help diagnose communication problems.