DataApex Clarity Control Module for Hitachi L6200/L6210 User Manual

This manual describes the installation, operation and troubleshooting of the Hitachi L6200/L6210 Control Module for controlling the Hitachi L6200/L6210 Liquid Chromatograph


DataApex Clarity Control Module for Hitachi L6200/L6210 User Manual - cover page
Brand
DataApex
Category
Application program
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
456 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
189E2ACA6E2611728330336E0B3377CE

About this manual

This manual describes the DataApex Clarity control module for Hitachi L6200 and L6210 pumps. The module enables direct control of the pump over a serial line from the Clarity environment, including the gradient program, and the instrument method is saved in the measured chromatograms.

It covers requirements, hardware wiring, communication settings, Clarity configuration, method setup, device monitoring, report setup and troubleshooting. Requirements include Clarity with the LC Control module (p/n A24), a free serial COM port and a serial straight DB9F-DB25M cable (p/n SKO3); the pump is controlled by RS232 communication with DIP Switch S1 on its Computer Board.

What's inside

  • Requirements
  • Installation procedure
  • Hardware - Wiring
  • Setting of the communication
  • Clarity Configuration
  • Method Setup - LC Gradient
  • Device Monitor
  • Hitachi L-6200 Setup
  • Report Setup
  • Troubleshooting

Specifications

Communication interfaceRS232
Baud rate4800
Data bits and stop bits8 databits + 2 stop bits
Required cableSerial straight DB9F-DB25M cable (p/n SKO3)
DIP switch locationS1 on Computer Board
Required Clarity moduleLC Control module (p/n A24)

Frequently asked questions

What does the Hitachi L6200/L6210 Control Module do?

It enables direct control of the Hitachi L6200 and L6210 pumps from the Clarity environment over a serial line, including the gradient program.

What are the communication settings for the Hitachi L6200/L6210?

The DIP switch settings include 8 databits + 2 stop bits and Baud rate 4800; DIP 6 is set OFF for RS232 mode.

What cable is needed for the serial connection?

The requirements list a serial straight DB9F-DB25M cable (p/n SKO3); the wiring section describes serial cross DB9F-DB25M wiring.

How can I record communication for troubleshooting?

Add or amend the COMMDRV.INI file in the Clarity installation directory to enable communication recording; the created TXT files help diagnose communication problems.