Getinge ED-FLOW, Automated Endoscope Reprocessor (AER) product temperature sensor may not work as intended, may provide incorrect temperature readings for washing and disinfection temperature during process, leading to product issues.
This document is an urgent Field Safety Notice (FSCN) from Getinge concerning the ED-FLOW Automated Endoscope Reprocessor (AER) and ED-FLOW SD AER. It addresses a potential hazard where the temperature sensor in the chamber may provide incorrect readings, displaying 'OK' even when temperatures do not meet validated biocidal activity requirements. The issue affects all ED-FLOW AER models manufactured from the start of 2010 until May 5, 2022.
A temperature sensor in the chamber may not work as intended and may provide incorrect temperature readings for washing and disinfection temperature during the process.
The difference between the Controlling value (CTRL) and the Independent value (I.V.) must not differ more than 2°C.
Getinge will release a new device software that provides an alarm when controlling temperature and Independent value temperature differ too much.