Getinge ED-FLOW AER Field Safety Notice Temperature Sensor Malfunction

Getinge ED-FLOW, Automated Endoscope Reprocessor (AER) product temperature sensor may not work as intended, may provide incorrect temperature readings for washing and disinfection temperature during process, leading to product issues.


Getinge ED-FLOW AER Field Safety Notice Temperature Sensor Malfunction - cover page
Brand
Getinge
Category
medical technology
Document type
Service Manual
Language
English
Pages
8
File format
PDF
File size
1.4 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
1C86938A3F8EFB480F3A5FC396D76704

About this manual

This document is an urgent Field Safety Notice (FSCN) from Getinge concerning the ED-FLOW Automated Endoscope Reprocessor (AER) and ED-FLOW SD AER. It addresses a potential hazard where the temperature sensor in the chamber may provide incorrect readings, displaying 'OK' even when temperatures do not meet validated biocidal activity requirements. The issue affects all ED-FLOW AER models manufactured from the start of 2010 until May 5, 2022.

What's inside

  • Product Issue
  • Affected Product
  • Resolution
  • The Issue
  • Action
  • Next Steps
  • Customer Response Form
  • Instruction Process Ticket
  • Critical Parameters
  • Recalibration

Frequently asked questions

What is the specific issue with the ED-FLOW AER?

A temperature sensor in the chamber may not work as intended and may provide incorrect temperature readings for washing and disinfection temperature during the process.

What is the maximum allowed difference between temperature sensors?

The difference between the Controlling value (CTRL) and the Independent value (I.V.) must not differ more than 2°C.

What corrective action is required for affected devices?

Getinge will release a new device software that provides an alarm when controlling temperature and Independent value temperature differ too much.