This manual introduces the operation methods of QUARTZ-35 and QUARTZ-50 dry block temperature calibrators, including installation, operation, maintenance and other contents.
This manual provides detailed instructions for operating and maintaining the DRUCK & TEMPERATUR QUARTZ-35 dry block temperature calibrator. It includes safety guidelines, setup procedures, temperature control functions, and maintenance steps. The device is designed for calibrating temperature sensors in both lab and field environments, with features like PID control, RS232 communication, and optional software for PC control. Safety precautions are emphasized, including proper grounding, avoiding thermal shock, and handling probes correctly to ensure accurate calibration and prevent damage.
The calibrator supports temperature ranges from -27°C to +150°C, with stability of ±0.02°C and resolution of 0.01°C. It includes a removable upper protection grid, internal cryostat with Peltier elements, and a flanged aluminum block with reduction inserts. Maintenance includes periodic recalibration and handling of the unit after sub-zero use to prevent moisture buildup.
| Operative Range | -27÷ +150°C @ 22°C |
|---|---|
| Stability | ±0.02°C |
| Resolution | 0,01/0,1°C |
| Reading Accuracy | ±0,15°C |
| Temperature Sensors | Pt 100 class A din43760 |
| Serial Communication | RS 232 |
| Heating Time | 15’ from 20 to 140°C (+ 6’ for stability) |
| Cooling Time | 12’ From 140 to 20°C (+ 6’ for stability) |
| Standard Block Size | Ø35 x 135mm |
| Power Supply | 230V 50Hz (100/115V by required) 50/60Hz |
| Power Consumption | 300VA |
| Weight | 10 Kg only calibrator; 17Kg with travelling-case |
| Microcontroller | 80C552 (family 80C51 CMOS) |
| Display | 2 lines 20ch x line (3,2x5,5) back lighting |
| A/D Converter | Σ-∆ 24 bits |
The operative range is -27°C to +150°C at 22°C ambient temperature.
The stability is ±0.02°C at -15°C and +150°C.