This document is about the maintenance manual of Sigma Delta Vaporizer device. The document mentions the maintenance content, maintenance process and maintenance personnel requirements of the device.
This service manual provides detailed instructions for servicing, repairing, and calibrating the Penlon Sigma Delta vaporizer. It includes safety warnings, service procedures, and parts lists. The vaporizer is designed for use in anesthesia machines and requires periodic service checks by trained engineers. Calibration must be performed using approved analyzers, and the unit must be serviced by authorized Penlon personnel or agents to ensure safety and accuracy.
The manual covers leak testing, calibration checks, overhaul procedures, and component replacement. It also includes warnings about handling anesthetic agents, storage conditions, and proper disposal of drained liquids. A parts list is provided with component descriptions and part numbers for reference during maintenance.
It is designed for incorporation in the fresh gas supply system of continuous flow anaesthetic machines.
The vaporizer must be purged at maximum output with 5 L/min flow of oxygen for 2 minutes, and the output checked with an analyser prior to use.
A Penlon trained engineer, an authorised Penlon agent, or a Penlon service engineer.