This manual mainly introduces the operation method of P-97 microelectrode drawing machine.
This manual provides step-by-step instructions for operating the Gowe-brother P-97 microelectrode puller, including setup, RAMPTEST calibration, parameter configuration, and maintenance. It details the front and rear panel controls, glass micropipette selection, and the mechanical and electronic systems involved in electrode fabrication. The guide also includes tips for optimizing performance with different glass types and troubleshooting common issues.
The device is designed for precise electrode fabrication using controlled heating, pulling, velocity, and cooling parameters. Users are instructed to perform a RAMPTEST to determine optimal heating values and adjust other parameters accordingly for desired electrode characteristics. This manual is written in Chinese.
| Heating Value (HEAT) | Determined by RAMPTEST; varies by platinum filament type |
|---|---|
| Pull Force (PULL) | Controls the force applied during pulling; higher values produce smaller, longer tips |
| Velocity (VELOCITY) | Controls the speed at which the pull rod moves; determines when the electrode breaks |
| Cooling Time (TIME) | Controls the duration of air cooling after heating |
The voltage converter should be set to 220V.
The recommended air pressure is 500.
The default setting for AIRMODE is not specified, but it can be switched to DELAY mode.