This document is about the management system of medical devices, including CT, MRI, X-ray machines, refrigerators, centrifuges, refrigerated centrifuges, electrocardiograms, and Holter electrocardiograms, including daily inspections and regular inspections.
This document outlines the inspection protocols for various medical devices used at the National Hospital Organization Kanazawa Medical Center. It specifies daily checks for equipment like CT scanners, MRI machines, and refrigerators, as well as annual or biannual maintenance by external vendors. Each device is listed with its manufacturer and model, and inspection frequency and procedures are clearly defined.
The manual also covers temperature monitoring for storage units, calibration requirements for thermometers, and specific inspection schedules for devices such as blood pressure monitors, scales, height meters, and electronic thermometers. It includes provisions for managing investigational devices and coordinating with sponsors for special cases. This manual is written in Japanese.
Daily inspection is performed on hospital operating days based on the inspection sheet determined by the radiology department and recorded.
Maintenance is performed four times per year by a contractor.
It is inspected once per year by a contractor, with the inspection result report kept on file.
Before use, it is checked to ensure power is on, the cuff inflates properly without air leakage, and there is no error display.