This reference guide introduces the methods for measuring fluorescence quantum yields and spectral reflectance using the FLS980 series integrating spheres
This manual explains how to prepare and use the FLS980 integrating sphere for measuring fluorescence quantum yields and spectral reflectance. It covers setup for liquid and solid samples, sample preparation, and instrumental factors affecting accuracy. The sphere features a 120 mm PTFE reflector with >99% reflectance from 400–1500 nm. Users must select correct correction files and ensure detector linearity for accurate results.
The document includes detailed steps for configuring the instrument, positioning mirrors and sample holders, and handling samples to avoid contamination. It also addresses special cases like low quantum yield samples and detector limitations, emphasizing careful planning and practice for reliable measurements.
| Sphere Diameter | 120 mm |
|---|---|
| Reflectance | >99% from 400–1500 nm, >95% from 250–2500 nm |
| Material | PTFE-like material |
| Detector Linearity Limit | 1 million photons per second |
Reflectance is >99% from 400–1500 nm and >95% from 250–2500 nm.
The recommended upper limit is about 1 million photons per second.
The sphere is made of PTFE-like material, and accessories like cuvette holders and powder trays are also made from PTFE.