Vacuum Technologies TriScroll 300 Series Tip Seal Replacement Manual

This document is a maintenance manual for the TriScroll 300 series dry scroll vacuum pump, which mainly introduces the process of replacing tip seals


Vacuum Technologies TriScroll 300 Series Tip Seal Replacement Manual - cover page
Brand
Vacuum Technologies
Document type
Service Manual
Language
English
Pages
31
File format
PDF
File size
822 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
F78FF32129E8DBB215146FF0C7D6774D

About this manual

This document is a draft Tip Seal Replacement Manual for the Vacuum Technologies TriScroll 300 Series Dry Scroll Vacuum Pump. It details the procedures for disassembling the pump to access the tip seals, cleaning the scrolls, and reassembling the unit. The manual also lists related maintenance kits and factory service options available from Vacuum Technologies.

What's inside

  • Preface
  • Safety Considerations
  • Maintenance and Tool Kits
  • Factory Service Options
  • Contacting Vacuum Technologies
  • Tip Seal Replacement
  • TriScroll 300 Disassembly
  • Scroll Cleaning
  • TriScroll 300 Reassembly
  • Putting the Pump Back into Service

Specifications

Weight26.4 kg (58 lbs)
Cool-down timeone to two hours
Run-in time24 hours

Frequently asked questions

When is tip seal replacement recommended?

Tip seal replacement is generally recommended when the pump base pressure has risen to an unacceptably high level for your application.

What lubricant is recommended for O-rings?

Vacuum Technologies recommends applying a small amount of Krytox GPL 224 grease and wiping the O-rings shiny dry.

Can the pump be put into service immediately after maintenance?

Yes, but 24 hours of run time is required before a base pressure of 10 mTorr can be achieved.