Introduces the technical parameters, installation methods, operation methods of Intell-Count™ Dual Channel Counting Scale IDC Series products
This manual covers operation and setup of Intell-Count IDC Series dual channel counting scales, including local and remote scale installation, key functions, display controls, and RS-232 communication. It details technical specifications for different models (IDC 3, 6, 15, 30) and calibration procedures. The scales support weighing, parts counting, memory accumulation, and PLU storage with unit weight and description entry. Error codes and troubleshooting are also included.
| Maximum Capacity | 3000 g, 6000 g, 15 kg, 30 kg |
|---|---|
| Readability | 0.05 g, 0.1 g, 0.2 g, 0.5 g |
| Tare Range | -3 kg, -6 kg, -10 kg, -30 kg |
| Interface | Bi-directional RS-232 Interface |
| Operating Temperature | 0°C - 40°C (32°F - 104°F) |
| Display | 3 x 6 digits LCD digital display |
| Power Supply | 9 VDC 800 mA from external power supply |
| Remote Scale Excitation Voltage | 5 VDC |
| Remote Scale Signal Range | 0-20 mv |
| Remote Scale Sensitivity | 0.02 µv/internal ADC count or better |
| Remote Scale Internal ADC Counts | 500,000 maximum at 10 mv input |
| Remote Scale Load | 87 ohm minimum, 4 X 350 ohm load cells |
| Remote Scale Connection | 4 wire connection to load cells plus shield |
| Remote Scale Maximum Cable Length | 20 ft - 6 meters |
| Remote Scale Termination | DB 9 subminiature plug on scale |
The IDC Series includes models with maximum capacities of 3000 g, 6000 g, 15 kg, and 30 kg.
The remote scale connects via a DB 9 pin d-subminiature plug with pins 1,2 for excitation+, 4,5 for excitation-, 7 for signal-, and 8 for signal+.
The RS-232 output uses ASCII code, 4800 Baud (600-9600 selectable), 8 data bits, no parity, and a 25-pin d-subminiature socket.
Error codes include Err 4 (Initial Zero greater than permissible value), Err 5 (Keyboard error), and Err 6 (A/D count not correct).