Thornton 200CR Communications Manual

This document introduces the serial communication interface of the Thornton Leading Pure Water Analytics 200CR instrument, including the settings of the RS232 and RS422 ports.


Thornton 200CR Communications Manual - cover page
Brand
Thornton
Category
Other Electronics
Document type
User Manual
Language
English
Pages
14
File format
PDF
File size
45 KB
Published
20 October, 2023
Updated
20 October, 2023
Model
200CR
MD5 checksum
84893EFA60BF36BE1389F4E3CE6E7741

About this manual

The Thornton 200CR instrument features a digital communication interface that can be configured as either an RS232 or RS422 port. Connections are made at the terminal block TB2 at the rear of the meter, and an external isolator is recommended to prevent ground loops. The interface allows the instrument to communicate with personal computers, programmable logic controllers, or printers.

What's inside

  • Wiring Connections
  • RS232 / RS422 Settings
  • Changing Baud Rate and Parity
  • Initialization Messages
  • Communicating with the 200CR
  • Response Messages
  • Automatic Data Output
  • Data Output Format
  • Command Set
  • Set Parameter Command

Specifications

Communication InterfacesRS232 or RS422
Default Baud Rate19,200
Data Bits8
Stop Bits1
Available Baud Rates19,200, 9600, 4800, 2400, and 1200
Data Output Interval1 second

Frequently asked questions

What are the available baud rates for the 200CR?

The available baud rates are 19,200, 9600, 4800, 2400, and 1200.

Can the number of data bits and stop bits be changed?

No, the number of data bits and stop bits cannot be changed.

What is the command to enable automatic data output?

Send the command 'B00' to enable the data output.

What is the response to a valid command?

A valid command is accepted with the response 'OK'.