ISIS Polaris Powder Diffractometer User Guide

This manual introduces the use of the ISIS Polaris powder diffractometer.


ISIS Polaris Powder Diffractometer User Guide - cover page
Brand
iSIS
Category
Other
Document type
User Manual
Language
English
Pages
31
File format
PDF
File size
449 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
5FB9818FDD5D4E5F3EE0CECC2C4DA846

About this manual

This guide explains how to perform powder diffraction experiments on the Polaris instrument at ISIS. It covers sample safety, loading, tank evacuation, data collection, and data assessment using software tools like GENIE and GSAS. The instrument features four detector banks (A, B, C, E) for high-intensity measurements with variable resolution. Safety protocols for handling irradiated samples are detailed, including storage and radiation monitoring requirements.

What's inside

  • Introduction
  • Performing an Experiment
  • Data Collection
  • Data Assessment
  • Sample Changer
  • Chopper Operation

Frequently asked questions

What is the maximum beam size at the sample position?

The maximum beam dimensions at the sample position are 40mm high by 20mm wide.

How many detectors are there in total on the POLARIS instrument?

POLARIS has a total of 434 detectors arranged into four separate banks.

What is the nominal pressure required to evacuate the sample tank?

The sample tank must be evacuated to a nominal pressure of approximately 0.1 mbar.