This manual describes the installation and operation of TC1 and TC2 tension cells, including inspection and handling, general safety, product description, mechanical installation, electrical installation, setup, maintenance and troubleshooting.
This manual provides instructions for the handling, storage, installation, and maintenance of Kistler-Morse TC1 and TC2 tension load cells. These sensors are designed for use in suspended in-process vessel weighing systems and are part of a full system that typically includes a junction box and PLC. The document covers functional checks, mechanical installation procedures for single, three, and four-point supports, and electrical wiring requirements.
| Functional Integrity | TC1: ± 150% of rated capacity; TC2: ± 150% of rated capacity |
|---|---|
| Mechanical Integrity | TC1: ± 300% of rated capacity; TC2: ± 500% of rated capacity |
| Excitation Voltage - Standard | 10 VDC |
| Excitation Voltage - Maximum | 15 VDC |
| Input Resistance | TC1: 350 ± 3.5 ohms; TC2: 800 ± 20 ohms |
| Nominal Output | TC1: 3 mV/V ± 0.5%; TC2: 1.0 mV/V ± 0.5% |
| Rated Full Scale Output | TC1: 30mV; TC2: 10mV |
| Environmental Protection Rating | IP67 |
| Cable Length | TC1: 10 ft (3 m); TC2: 16 ft (5 m) |
Both the TC1 and TC2 have a functional integrity of ± 150% of rated capacity.
The standard excitation voltage is 10 VDC.
The TC1 has an input resistance of 350 ± 3.5 ohms, while the TC2 is 800 ± 20 ohms.
Both models have an environmental protection rating of IP67.