This manual describes the operation and maintenance of the Nuaire LabGard® ES HD Energy Saver Class II, Type A2 Laminar Flow Biosafety Cabinet
The Nuaire LabGard ES HD Energy Saver Class II, Type A2 Laminar Flow Biosafety Cabinet (Models NU-581-400E/500E/600E) is a bench or console model designed for biological research and hazardous drug preparation. It utilizes an Energy Saver DC ECM motor with digital dual thermistor airflow sensors to provide constant volume control, compensating for filter loading and line voltage variances. The cabinet meets the requirements of NSF/ANSI 49, EN 12469, and DIN 12980:2016-10 standards.
| Class | Class II, Type A2 |
|---|---|
| Inflow Velocity | 105 FPM (.53 m/s) |
| HEPA Filter Efficiency | -99.995% Eff. On MPPS |
| Voltage | 230 Volts, AC 50/60 Hz |
| Sound Pressure Level | Not to exceed 58 dbA (4ft model) |
The cabinet conforms to NSF/ANSI 49, EN 12469, and DIN 12980:2016-10 standards.
Yes, Type A2 cabinets may be used for work with volatile chemicals if permitted by a chemical risk assessment, but must be connected to an external exhaust system with a canopy transition.