This manual covers use and maintenance of the MVE Automatic LN2 Supply Switch
The MVE Automatic LN2 Supply Switch is an automated backup device designed to switch from a primary liquid nitrogen supply to a backup supply when the primary is depleted. It is intended for use by experienced personnel and eliminates the need for an electrical signal from the freezer to indicate LN2 demand, thereby reducing electrical noise issues. The unit utilizes pressure switches to detect supply status and demand, activating solenoid valves to maintain supply to the connected device.
| Operating Pressure Range | 22psi to 35psi |
|---|---|
| Empty Indicator Threshold | below 10 psi |
| Expansion Ratio | 1 to 700 |
| Liquid Density | 1.782 lb./liter |
This Technical Manual is intended for use by experienced personnel only.
The switch uses downstream pressure in the transfer line to detect demand, eliminating the need for an electrical signal from the freezer.