This operation manual provides instructions for operating the Locator and Locator Plus cryobiological storage vessels and liquid nitrogen level monitors.
This manual provides operating instructions and safety guidelines for Barnstead|Thermolyne's Locator® and Locator® Plus cryogenic storage vessels and liquid nitrogen level monitors. It covers filling procedures, measuring liquid nitrogen levels, storing samples in vapor phase, and safety precautions to prevent vacuum failure. The document also includes part numbers and specifications for different vessel models and accessories.
The manual emphasizes safety, including handling liquid nitrogen with care, avoiding overfilling, and protecting the vacuum seal. It provides detailed instructions for filling new or warm vessels, measuring liquid nitrogen levels, and storing samples in vapor phase. Replacement parts and accessories are listed with model-specific part numbers.
| Height (Including lid) | 37.5" to 40" (95.25 cm to 101.6 cm) |
|---|---|
| Diameter | 22" (55.8 cm) |
| Vessel Volume | 110L to 165L |
| LN2 Capacity (Liquid Phase) | 90L to 134L |
| LN2 Capacity (Vapor Phase) | 25L to 39L |
| Evaporation Rate (Liquid Phase) | 2916 L/Day to 4860 L/Day |
| Level Monitor Voltage | 16 volts (DC) |
The liquid nitrogen level should never be above 20 inches (50 cm) for LOCATOR 4 and LOCATOR 8.
The liquid nitrogen level should never be above 10 inches (25 cm) for LOCATOR Jr. and 12 inches (30 cm) for LOCATOR Jr. Plus.
This can cause vacuum failure. Do not spill liquid nitrogen on the vacuum port.