This manual contains warnings and safety rules that must be observed by the user to ensure safe operation of the KEW SNAP 2432/2433 series leakage current tester and to retain it in safe condition
This document is the instruction manual for the Kyoritsu SNAP 2432 and 2433 leakage current testers. It provides essential safety warnings, operating instructions, and feature descriptions for these digital clamp-on instruments designed for measuring AC leakage current.
The manual details how to use the frequency selector button to filter high-frequency noise from inverters, how to perform peak current measurements, and explains the auto-power-off function. It also outlines the measurement categories (CAT I, II, III, IV) and safety standards (IEC 61010) applicable to the device.
| Display Height | 18mm |
|---|---|
| Dynamic Range | 4000 counts |
| Peak Hold Resolution | 10msec |
| Auto-Power-Off Time | 10 minutes |
| High-Cut Filter Cutoff | approx. 160Hz |
Press the Peak Hold Button to set the instrument to peak measurement mode, indicated by a 'P' on the display.
It permits measurement of currents of fundamental frequencies as well as currents of high frequencies generated by equipment like inverters.