This is a manual for EX50 Excimer laser
This service manual provides detailed instructions for operating, maintaining, and troubleshooting the GAM LASER EX50 excimer laser. It covers laser specifications, water-cooled operation, gas handling, chamber exchange, electronics, and troubleshooting common issues. Safety guidelines and operational limits are emphasized to prevent damage. The manual applies to models EX50/250, EX50/500, EX50/750, and EX50/1000.
The document includes sections on optics cleaning, gas refills, thyratron adjustment, and interlock systems. It also outlines procedures for replacing components, handling gas mixtures, and ensuring proper ventilation and cooling. The manual is intended for authorized personnel and includes warnings about unauthorized operation or maintenance.
| Wavelength | 157 nm, 193 nm, 248 nm, 308 nm, 351 nm |
|---|---|
| Repetition Rate | 250/500/750/1000/2000 Hz |
| Cooling Type | Air-cooled or Water-cooled |
| Thyratron Warm-up Time | 5-8 minutes |
Do not exceed a duty cycle of approximately 80% in laser operation over the long term for air cooled units, (1 hour) at the maximum repetition rate of 500Hz. For shorter periods of time the laser should be operated for no greater than 20 minutes at maximum repetition rate. Water cooled units can be operated at maximum duty cycle.
Do not exceed a maximum high voltage input of 16 kV. Hardware protection will prevent the high voltage exceeding 16.6kV.
Service should be scheduled after every 6 months of operation. At scheduled service, optics should be cleaned or replaced, internal optical elements associated with the energy detector should be cleaned, and the thyratron filament voltage tested and reset if required.