This manual introduces the correct use of the IWAKI AMERICA CFD BELLOWS PUMP, including unpacking and inspection, specifications, operation methods, maintenance and inspection, and troubleshooting.
This manual provides detailed instructions for operating, maintaining, and troubleshooting IWAKI CFD bellows pumps used for metering high purity chemicals. It covers handling, installation, wiring, operation with controllers, and maintenance procedures. Safety precautions are emphasized, especially when handling hazardous chemicals. The pump can be controlled by computers, PLCs, or Iwaki APD controllers, with specific wiring and valve requirements outlined.
| Max. discharge per shot | 0.33 oz (10ml) |
|---|---|
| Max. pressure | 7.11 PSI (0.5 kgf/cm2) |
| Max. stroke spm | 30 |
| Supply air pressure | 21-43 PSI (1.5-3.0 kgf/cm2) |
| Temperature range of liquid handled | 68-140°F (20-60°C) |
| Air consumption | .006 Ft3/shot (0.18 Nl/shot) |
| Wet end material | PTFE, PFA |
| Pump connection port diameter | 1/4" OD tube (∅ 6.35 x ∅ 4.35mm) |
| Input voltage, Sensors | 5 VDC |
| Output voltage, Sensors | 0 – 5 VDC |
| Weight | 3.3 lbs (1.5 kg) |
The pump can be operated directly by host computers, PLCs, or with Iwaki APD-1 or APD-3 controllers.
The pump is designed for use in metering high purity chemicals, including considerably dangerous liquids such as strong acids.
The maximum discharge per shot capability for the CFD-8T model is 0.33 oz (10ml).
The pump requires a supply air pressure between 21 and 43 PSI (1.5-3.0 kgf/cm2).