This manual introduces the use of KOBOLD KAL-K (KAL-4000 series) thermal flow switch, including product specifications, mechanical installation, electrical connections, operation, calibration, flow setpoint adjustment, setting output transistor type, troubleshooting, applications, maintenance, damaged equipment arrival and need help.
The KOBOLD KAL-K (KAL-4000 Series) is a thermal flow switch designed for monitoring and controlling moderate flowrates of non-viscous or dirty liquids. It operates on the thermal dispersion principle, using a heated probe to detect flow velocity and compare it against a user-selected setpoint. The device features a microprocessor-controlled design that allows for simple calibration and setup, including zero-flow alignment and trend indicator span adjustment.
| Measuring Range | 0.13-6.6 feet/sec. (0.05 - 2 meter/sec.) |
|---|---|
| Fitting Size | 1/2" NPT |
| Maximum Pressure | 1450 PSIG |
| Maximum Temperature | 250°F |
| Supply Power | 24 VDC ± 10%, 300 mA Max. |
It uses the proven thermal dispersion principle.
It is recommended to install the unit in the upper hemisphere of the pipe.
The factory preset span corresponds to a water flow velocity of 2 meters/second.