This document describes the operating manual of HY-ALERTA™ 2620 Explosion Proof Area Hydrogen Monitor.
This document provides detailed instructions for operating, installing, and maintaining the HY-OPTIMA™ 2620 hydrogen monitor in explosion-proof areas. It includes wiring diagrams, calibration procedures, relay configurations, and serial communication commands. The monitor is designed for hydrogen detection with analog and digital outputs, and features certifications for use in hazardous locations. Maintenance and troubleshooting are also covered, including cleaning and field calibration steps.
| Sample Gas Flow Rate | 0 to 50 slpm (1slpm is recommended) |
|---|---|
| Input Power | 90VAC to 240VAC, 15W, 50Hz to 60Hz |
| Analog Output | Isolated and self-powered 0-20mA, maximum load impedance: 650 Ω |
| Relay Contacts | Three programmable SPDT relays with both normally open (N.O.) and normally closed (N.C.) contacts. 5A, 240VAC MAX |
| Serial Communication | RS232 or RS422 19200 bps, No parity, 8 bit data, 1 stop bit, No flow control (preferred) or Xon/Xoff |
| Dimensions | See following figures |
| Weight | 2.7kg (5.9lbs) |
| Temperature Sensor Ambient | -20°C to +55°C |
| Storage Temperature | -40 to +80 °C |
| Accuracy | ± (0.03 x indication + 0.2) percent hydrogen by volume (serial port output only) |
Ambient temperature range is -20°C to +55°C for operation, and -40°C to +80°C for storage.
It requires 90VAC to 240VAC, 15W, 50Hz to 60Hz.
Three programmable SPDT relays with both normally open and normally closed contacts, rated at 5A, 240VAC MAX.
Isolated and self-powered 0-20mA, with a maximum load impedance of 650 Ω.