The Teledyne Monitor Labs LaserHawk 360 is a laser-based particulate monitor designed to continuously monitor dust loading in combustion exhaust gas streams. Its electrical output is proportional to light scattered back from particles in the gas stream and may be conditioned by the Enhanced Remote Panel or a user-supplied device.

The manual describes installation, operation, calibration and routine maintenance. It covers three models: the LaserHawk 360 and 360DI with a minimum particulate level of 20 mg/M3, and the extra-sensitive LaserHawk 360ES with a minimum particulate level of 5 mg/M3. It also covers the optical head assembly, protective purge air system, Enhanced Remote Panel, calibration kit and laser safety.

What's inside

  • System Overview
  • Specifications
  • Optional Equipment
  • Optical Head Assembly
  • Enhanced Remote Panel
  • Purge System Overview
  • Calibration Kit Overview
  • Calibration & Checkout
  • Installation
  • Maintenance

Specifications

Optical measurement techniqueOptical back-scatter of light from a red laser
Spectral response655 nanometers nominal, +/- 20 nanometers
Resolution0.5 mg per actual m3
Response timeLess than 10 seconds to 95% of change
Optical head power requirements85-245 VAC, 47-63 Hz, single phase, 30 VA maximum
Optical head weight28 lbs. (12.7 kg)

Frequently asked questions

What is the minimum particulate level for the LaserHawk 360ES?

The LaserHawk 360ES has a minimum particulate level of 5 mg/M3.

What optical measurement technique does the monitor use?

It uses optical back-scatter of light from a red laser.

What is the response time of the LaserHawk 360?

The response time to 95% of a change is less than 10 seconds.

What is the minimum particulate level for the standard LaserHawk 360?

The standard LaserHawk 360 has a minimum particulate level of 20 mg/M3.