This document is the user manual of Kanmed Warming Cabinet GE-2-1370, which mainly introduces the functions, usage methods, installation methods and maintenance methods of the product.
The Kanmed Warming Cabinet GE-2-1370 is a stainless steel medical device designed to warm gel pads, blankets, and fluids up to 42°C. It features a heating element, fan, and thermostat for even air circulation, with double-glassed doors to reduce heat loss and noise. The cabinet is intended for continuous use in hospital settings and must be secured to a wall to prevent tipping.
| Voltage | 220-240 Volt AC |
|---|---|
| Power | max. 1200 Watt |
| Frequency | 50 -60 Hz |
| Fuses | T 6,3A L250V |
| Height | 175 cm |
| Width | 66 cm |
| Depth | 60 cm |
| Working temperature | 35 °C – 42°C |
| Over temperature protection | 45 °C |
| Accuracy | ± 1°C |
| Hysteresis | 4 °C max |
| Warming capacity | 10 gel pads 3 hours |
The cabinet is primarily intended for warming contents up to a temperature of 42°C.
The maximum load per shelf is 8 kilos.
Temperature control and calibration must be performed by a qualified technician only.
The thermostat T2 takes over control, and a red lamp on the front panel lights up to indicate a malfunction.