About this manual

The Kanmed Warming Cabinet GE-2-1370 is a stainless steel medical device designed to warm gel pads, blankets, and fluids up to 42°C. It features a heating element, fan, and thermostat for even air circulation, with double-glassed doors to reduce heat loss and noise. The cabinet is intended for continuous use in hospital settings and must be secured to a wall to prevent tipping.

What's inside

  • Safety Instructions
  • General Description
  • Description of Function
  • Technical Data
  • Unpacking and Packing List
  • Installation
  • Using the Cabinet
  • Temperature Adjustment
  • Maintenance and Troubleshooting
  • Accessories and Spare Parts

Specifications

Voltage220-240 Volt AC
Powermax. 1200 Watt
Frequency50 -60 Hz
FusesT 6,3A L250V
Height175 cm
Width66 cm
Depth60 cm
Working temperature35 °C – 42°C
Over temperature protection45 °C
Accuracy± 1°C
Hysteresis4 °C max
Warming capacity10 gel pads 3 hours

Frequently asked questions

What is the maximum temperature the cabinet can reach?

The cabinet is primarily intended for warming contents up to a temperature of 42°C.

What is the maximum load per shelf?

The maximum load per shelf is 8 kilos.

How often should the temperature be calibrated?

Temperature control and calibration must be performed by a qualified technician only.

What happens if the temperature exceeds 45°C?

The thermostat T2 takes over control, and a red lamp on the front panel lights up to indicate a malfunction.