EC35 gas sensor can continuously detect gas concentrations and warn of toxic gases and oxygen deficiency in ambient air.
The EC35 is an intelligent gas transmitter for continuous detection of toxic gases and oxygen deficiency. It features plug-in smart sensors, scalable detection ranges, and a 4-20mA output. The manual covers mounting, electrical connections, operation modes, calibration, and troubleshooting. It also includes safety warnings and service instructions for maintaining accuracy and safety.
The device is designed for use in hazardous environments and includes remote control options. Calibration and zero-point adjustments are automated via the AutoCal function. Sensor life is indicated, and the unit requires periodic maintenance to ensure accurate readings.
| Output | 4-20mA |
|---|---|
| Explosion Protection | IP 66 |
| Sensor Type | Electrochemical cells |
| Current Consumption | Approx. 25mA at 24V DC |
Press ZERO for at least 3 seconds to activate the AutoCal program. The display will show the measurement value and ZERO alternately. If the measurement value remains constant, it is overwritten by the zero value. The AutoCal program stops automatically.
The EC35 indicates the expired lifetime by reading CHNG SENS and measurement value alternately. The sensor should be replaced as soon as possible.
Allow a warm-up time of 30 minutes before checking the zeropoint. After this time, the display should have stabilized to read "000".
Press SPAN to display: Unit of measurement, Gas, Detection range, and Calibration gas concentration.