GfG Instrumentation, Inc. EC35 Gas Transmitter Operation Manual

EC35 gas sensor can continuously detect gas concentrations and warn of toxic gases and oxygen deficiency in ambient air.


GfG Instrumentation, Inc. EC35 Gas Transmitter Operation Manual - cover page
Brand
GfG Instrumentation, Inc.
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
160 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
A949182C479822F8AF3DAA2081B6CE42

About this manual

The EC35 is an intelligent gas transmitter for continuous detection of toxic gases and oxygen deficiency. It features plug-in smart sensors, scalable detection ranges, and a 4-20mA output. The manual covers mounting, electrical connections, operation modes, calibration, and troubleshooting. It also includes safety warnings and service instructions for maintaining accuracy and safety.

The device is designed for use in hazardous environments and includes remote control options. Calibration and zero-point adjustments are automated via the AutoCal function. Sensor life is indicated, and the unit requires periodic maintenance to ensure accurate readings.

What's inside

  • General Description
  • Detection Principle
  • For Your Safety
  • Design
  • Mounting
  • Installation of Electrical Connections
  • Putting in Operation
  • Detection Mode
  • Check of Display and Parameter
  • Sensor Life

Specifications

Output4-20mA
Explosion ProtectionIP 66
Sensor TypeElectrochemical cells
Current ConsumptionApprox. 25mA at 24V DC

Frequently asked questions

How do I calibrate the zero point of the EC35?

Press ZERO for at least 3 seconds to activate the AutoCal program. The display will show the measurement value and ZERO alternately. If the measurement value remains constant, it is overwritten by the zero value. The AutoCal program stops automatically.

What happens when the sensor life expires?

The EC35 indicates the expired lifetime by reading CHNG SENS and measurement value alternately. The sensor should be replaced as soon as possible.

How long does the EC35 need to warm up before checking the zeropoint?

Allow a warm-up time of 30 minutes before checking the zeropoint. After this time, the display should have stabilized to read "000".

What is the detection range display format?

Press SPAN to display: Unit of measurement, Gas, Detection range, and Calibration gas concentration.