Steam Sterilizer Service Manual
This service manual provides detailed instructions for maintaining and repairing Trans Medikal's SMART and SMART-S steam sterilizers. It covers chamber capacities, jacket and generator specifications, and troubleshooting for components like steam lines, vacuum lines, and sensors. The manual also includes safety guidelines and maintenance procedures for ensuring proper operation and longevity of the equipment.
The sterilizer is classified as Class IIb medical equipment under MDD 93/42/EEC, with a 15-year product life cycle. It is constructed with AISI 316L stainless steel for durability and corrosion resistance, and includes technical specifications for operating pressures and safety vents.
| Chamber Material | AISI 316L stainless steel (X2CrNiMo17-12-2) |
|---|---|
| Chamber Sheet Thickness | 6 mm |
| Chamber Operating Pressure | 2.2 Bar |
| Chamber Safety Valve Pressure | 3 Bar |
| Jacket Material | AISI 316L stainless steel |
| Jacket Sheet Thickness | 3 mm |
| Jacket Operating Pressure | 1.2 bars |
| Generator Material | AISI 316L stainless steel (X2CrNiMo17-12-2) |
| Generator Sheet Thickness | 3 mm |
| Generator Operating Pressure | 2.5 bar |
| Generator Safety Vent Pressure | 4 bar |
| Heater Performance | 10Kw or 15 Kw |
The life cycle of the Steam Sterilizer (Autoclave) is 15 (fifteen) years, as determined by the Medical Device Directive (MDD-93/42/EEC).
The chamber is produced of AISI 316L stainless steel (X2CrNiMo17-12-2).
The generator operating pressure is 2.5 bar.
The generator safety vent releasing pressure is 4 bar.