This manual describes the basic operations of the Geotech BIOGAS 5000 gas analyser, including the composition of the equipment, safety use, installation and maintenance.
The BIOGAS 5000 is a portable gas analyser designed for measuring gases from landfill sites and other sources. It enables consistent data collection for improved analysis and accurate reporting while helping to check the digester process efficiency. The device complies with Part 15 of the FCC Rules and is certified for hazardous area classification II 2G Ex ib IIA T1 Gb.
| Hazardous Area Classification | II 2G Ex ib IIA T1 Gb |
|---|---|
| Operating Temperature Range | -10ºC to +50ºC |
| Maximum Oxygen Limit | 21% |
| Connector C Specifications | Uo=10V, Io=5mA, Po=50mW |
The equipment is only certified for use in ambient temperatures in the range -10ºC to +50ºC.
No, the equipment must not be used in an atmosphere of greater than 21% oxygen.
Only battery pack part numbers 20087 or 2011113 are permitted as replacements.