Instruction manual for 2020we/wi 1970-EPA 1970-ISO turbidity meter

This manual is the instruction manual for the 2020we/wi 1970-EPA 1970-ISO turbidity meter, which mainly introduces the use method and precautions of the turbidity meter.


Instruction manual for 2020we/wi 1970-EPA 1970-ISO turbidity meter - cover page
Brand
LaMotte
Category
Other Electronics
Document type
User Manual
Language
English
Pages
52
File format
PDF
File size
309 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
BABA61EC2F9AE78D5042A9B9BFC42CEB

About this manual

The LaMotte 2020we/wi is a nephelometer designed to measure turbidity in water samples. The 2020we model meets EPA specifications using a tungsten lamp and measures in NTU, while the 2020wi model meets ISO specifications using an LED and measures in FNU. The device utilizes both attenuation and nephelometric methods to ensure accurate readings across a wide range of turbidity levels.

What's inside

  • Introduction
  • Taking Turbidity Water Samples
  • Options & Setup
  • Data Logging
  • Calibration & Analysis
  • Troubleshooting Guide
  • General Operating Information
  • Computer Connection
  • Battery Operation
  • Maintenance

Specifications

Light Source (EPA)tungsten lamp with a color temperature of 2,200–3,000 K
Light Source (ISO)light emitting diode (LED) with a wavelength of 860 ± 30 nm
EPA Detector Angle180°
ISO Detector Angle90°

Frequently asked questions

What is the difference between the 2020we and 2020wi models?

The 2020we meets EPA specifications and measures in NTU, while the 2020wi meets ISO specifications and measures in FNU.

Can I use Stablcal standards to calibrate the meter?

No, Stablcal standards below 50 NTU should not be used to calibrate the 2020we/wi.

What are the default factory settings for the meter?

The default settings include Averaging Disabled, Turbidity Units NTU, Calibration Formazin, and Power Save set to 5 minutes.